Devices Supported: |
Sartorius LP 6200 S Balance |
Device
Manufacturer: |
Sartorius AG Goettingen, Germany TEL: (551) 308-0
FAX: (551) 308-289 |
Profile Name: |
Sartorius_LP6200S.PRO |
Profile Built For: |
UOS |
Platform: |
Windows NT4.0 / 2000 / XP |
Interface Type: |
RS-232 |
Max Devices
Supported: |
1 |
Protocol: |
Sartorius LP 6200 S Balance Protocol |
Comment: |
ULP is comprised of the Universal OPC
Server (UOS) and the Sartorius_LP6200S.pro device profile. The ULP
mnemonic is used to identify this unique combination. UOS will also support many other
devices that use serial communications. Please refer to our web site for the current list. |
|
UOS is a stand-alone OPC Server that can be used
with any OPC compliant Client such as Intellutions FIX® or FIX Dynamics®
or OAs Enterprise Automation software. UOS has many advanced features like:
DataView, DataScope, Serial and TCP/IP communications support, and multiple
"different" device support. Please refer to our web site for more information on
UOS capabilities. |
Contact: |
Dimension Software, Inc.
1536 St. Clair Road, Taylorsville, NC 28681, USA
Voice: (828) 635-7189, Fax: (828) 625-5319,
E-Mail: ,
Web: http://dimension-software.com |
References: |
UOS (Universal OPC Server)
Electronic Documentation. This includes context sensitive documentation on UOS
configuration and device dependent profile documentation. |
|
Sartorius Master pro Series LP Models Electronic
Precision Balances Operating Instructions. Technical Support: 1-800-635-2906 |
Profile
Command Group Name |
Description |
PrintRaw |
Use this command group to Print the raw data weight displayed
on the LP6200S Balance. The balance Level 7 (Print for Application) MUST be set to 7 2 1. SARTORIUS
LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: P
Description: Read/Print weight and units displayed on balance display (Format 1
command) |
PrintGLP |
Use this command group to Print the GLP data weight displayed
on the LP6200S Balance. The balance Level 7 (Print for Application) MUST be set to 7 2 2. SARTORIUS
LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: P
Description: Read/Print GLP weight and units displayed on balance (Format 1 command) |
AutoPrint |
Use this command group to read unsolicited raw data weight
displayed on the LP6200S Balance. The balance Level 6 (Print Manual/Automatic) MUST be set
to 614, 615, or 616. The balance Level 7 (Print for Application) MUST be set to 7 2 1. SARTORIUS
LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: None
Description: Unsolicited Weight Data |
Tare |
Use this command group to Tare the LP6200S Balance. Using
this TARE command performs the same action as pressing the TARE button on the balance. SARTORIUS
LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: T
Description: Tare Balance (Format 1 command) |
Restart |
Use this command group to Restart the LP 6200 S Balance. SARTORIUS
LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: S
Description: Restart Balance (Format 1 command) |
Weighing_1 |
Use this command group to Set Weighing Mode 1 (min
vib) on the LP 6200 S Balance. SARTORIUS LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: K
Description: Set Weighing Mode 1 (Format 1 command) |
Weighing_2 |
Use this command group to Set Weighing Mode 2
(norm vib) on the LP 6200 S Balance. SARTORIUS LP6200S COMMANDS USED IN THIS COMMAND
GROUP:
Command: L
Description: Set Weighing Mode 2 (Format 1 command) |
Weighing_3 |
Use this command group to Set Weighing Mode 3
(strong vib) on the LP 6200 S Balance. SARTORIUS LP6200S COMMANDS USED IN THIS COMMAND
GROUP:
Command: M
Description: Set Weighing Mode 3 (Format 1 command) |
Weighin_4 |
Use this command group to Set Weighing Mode 4
(extr vib) on the LP 6200 S Balance. SARTORIUS LP6200S COMMANDS USED IN THIS COMMAND
GROUP:
Command: N
Description: Set Weighing Mode 4 (Format 1 command) |
Tare_2 |
Use this command group to Tare the LP6200S
Balance. Using this TARE command Taring is not possible when the gross weight is less than
or equal to zero. SARTORIUS LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: f4
Description: Tare Balance (Format 2 command) |
Zero |
Use this command group to Zero the LP6200S
Balance. SARTORIUS LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: f3
Description: Zero Balance (Format 2 command) |
PrintModel |
Use this command group to Print the Balance Model
for the LP 6200 S Balance. SARTORIUS LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: x1
Description: Print Balance Model (Format 2 command) |
PrintSerial |
Use this command group to Print the Weighing Cell
Serial Number for the LP 6200 S Balance. SARTORIUS LP6200S COMMANDS USED IN THIS
COMMAND GROUP:
Command: x3
Description: Print Weighing Cell Serial Number (Format 2 command) |
R/W_ID |
Use this command group to Print and set the ID
Number for the LP6200S Balance. SARTORIUS LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: x5
Description: Read/Print ID number (Format 2 command)
Command: z5
Description: Set ID number (Format 3 command) |
R/W_Weight |
Use this command group to Print and set the
Weight Set Number for the LP6200S Balance. SARTORIUS LP6200S COMMANDS USED IN THIS
COMMAND GROUP:
Command: x6
Description: Read/Print Weight set number (Format 2 command)
Command: z6
Description: Set Weight Set number (Format 3 command) |
R/W_Lot |
Use this command group to Print and set the Lot
Number for the LP6200S Balance. SARTORIUS LP6200S COMMANDS USED IN THIS COMMAND GROUP:
Command: x7
Description: Read/Print Lot number (Format 2 command)
Command: z7
Description: Set Lot number (Format 3 command) |
TextInput |
Use this command group to Write a maximum of 8
text characters to the LP6200S Balance display. User must clear the display using the
"F" function key or the "F" command group. SARTORIUS LP6200S
COMMANDS USED IN THIS COMMAND GROUP:
Command: t
Description: Text Input In Display (Format 3 command) |
F_Key |
Use this command group create the same action as
pressing the "F" Function Key for the LP 6200 S Balance. SARTORIUS LP6200S
COMMANDS USED IN THIS COMMAND GROUP:
Command: f2
Description: F Function Keyl (Format 2 command) |
Signal Conditioning
Name: |
Description: |
NONE |
No Scaling |
LIN (Unsigned) |
Scales 16-bit unsigned values to the database block's EGU
range. |
LIN (Signed) |
Scales 16-bit signed values to the database block's EGU
range. |
3BCD |
Scales 3-digit Binary Coded Decimal values to the database
block's EGU range. |
4BCD |
Scales 4-digit Binary Coded Decimal values to the database
block's EGU range. |
8BN |
Scales 8-bit binary values to the database block's EGU
range. Ignores the most significant byte. No Alarming on Out of Range. |
8AL |
Scales 8-bit binary values to the database block's EGU
range. Ignores the most significant byte. Alarming on Out of Range. |
12BN |
Scales 12-bit binary values to the database block's EGU
range. Ignores the most significant nibble (4-bits). No Alarming on Out of
Range. |
12AL |
Scales 12-bit binary values to the database block's EGU
range. Ignores the most significant nibble (4-bits). Alarming on Out of
Range. |
13BN |
Scales 13-bit binary values to the database block's EGU
range. Ignores the most significant 3 bits. No Alarming on Out of Range. |
13AL |
Scales 13-bit binary values to the database block's EGU
range. Ignores the most significant 3 bits. Alarming on Out of Range. |
15BN |
Scales 15-bit binary values to the database block's EGU
range. Ignores the most significant bit. No Alarming on Out of Range. |
15AL |
Scales 15-bit binary values to the database block's EGU
range. Ignores the most significant bit. Alarming on Out of Range. |